Open original FDA record ↗Recall record
- FDA identifier
- Z-2881-2026
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Aug 4, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Source facts
- Action
- On June 26, 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were informed of the issue and provided with revised Safety Data Sheets (SDS) specific to the impacted lots for handling and disposal information.
- Initiated
- 2026-06-26
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.
- Recall number
- Z-2881-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221448
- FDA classified
- 2026-08-04
- Recall event number
- 99381