Skip to content

Public FDA record

Z-2881-2026: Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

Recall record Z-2881-2026 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2881-2026
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Aug 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

Source facts

Action
On June 26, 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were informed of the issue and provided with revised Safety Data Sheets (SDS) specific to the impacted lots for handling and disposal information.
Initiated
2026-06-26
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.
Recall number
Z-2881-2026
Recall status
Open, Classified
CDRH record ID
221448
FDA classified
2026-08-04
Recall event number
99381

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.