Z-2882-2020: BD Alaris System; Alaris Syringe Module Model 8110; BD Alaris System; Alaris PCA Module Model 8120; B

Recall record Z-2882-2020 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2882-2020
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Sep 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion.

Source facts

Action
The recall letter was sent via FedEx, 2nd day air, on August 4, 2020. On December 16, 2022, CareFusion expanded the recall to include additional affected customers. The notification was mailed via FedEx, 2nd day Air. Action for Clinical Users: Ensure that the displayed syringe manufacturer and syringe size correctly identify the installed syringe. If the PC unit displays an incorrect syringe type and/or size, remove the device from service and send to Biomedical Engineering for repair. The biomedical engineer should: - Conduct a barrel clamp accuracy test as outlined in the "Preventive Maintenance" section of the BD Alaris" System Maintenance - Software User Manual and replace syringe barrel clamp mechanism if required. - Contact BD at 1-800-482-4822 to order a Syringe/PCA Sizer Sensor Replacement Kit at no charge, or - Contact the BD Recall Support Center at 888-562-6018 to send the module to the BD Service Depot for repair at no charge. Actions by BD: BD is redesigning the syringe barrel clamp mechanism as a quality improvement.
Initiated
2020-08-04
Root cause
Component design/selection
Distribution
US - AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY OUS - AE, AU, BE, BH, CA, CH, CN, IN, KW, MX, MY, NL, NZ, PH, QA, SA, TW and ZA
Recall number
Z-2882-2020
Recall status
Open, Classified
CDRH record ID
182421
FDA classified
2020-09-04
Recall event number
86005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗