Skip to content

Public FDA record

Z-2885-2026: Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Numbe

Recall record Z-2885-2026 for United Orthopedic Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2885-2026
Company in FDA record
United Orthopedic Corp.
FDA classification date
Aug 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

Source facts

Action
On 03/11/2026, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that due the fact that the PSA Box Trial cannot be properly assembled with the Hinge Trial in the affected devices because of incomplete processing of the screw components, which may result in interference during assembly with the corresponding device. On 05/20/2026, the firm emailed an updated "URGENT: MEDCIAL DEVICE RECALL" Letter Customers are instructed: 1. Examine your inventory immediately to locate the affected product whose quantities and lot numbers are listed above. Please note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in your territory for immediate return. 3. Customer Service will be contacting you regarding replacements. If you still have the product, you are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if your firm has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For Question - contact UOC Sr. Director of Sales Operations at 949.328.3366.
Initiated
2026-05-20
Root cause
Process design
Distribution
US Nationwide distribution in the states of Texas and Michigan.
Recall number
Z-2885-2026
Recall status
Open, Classified
CDRH record ID
220571
FDA classified
2026-08-05
Recall event number
99026

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗