The procedure side indicated in the top header of the Shoulder Case Planning Report incorrectly indicated the procedure side as "Left" for surgical plans for a "right" procedure side.
Source facts
Action
The firm, materialise, sent Recall letters dated 17 July 2020 informing customers of the issue. This letter informs the customer that Materialise has corrected the Shoulder Case Planning Report in the SurgiCase online system.
If you have any questions, contact Global Quality and Regulatory Manager Medical, at 734-259-7017 or email: Jenny.Jones@materialise.com.
Initiated
2020-07-17
Root cause
Software Design Change
Terminated
2020-12-23
Distribution
Nationwide distribution to: US including PR and International to UK and Australia.
Recall number
Z-2886-2020
Recall status
Terminated
CDRH record ID
182679
FDA record created
2020-08-26
Recall event number
86113