- Recall classification
- Class II
Reason stated in the source
Two analgesic gas delivery system oxygen/nitrous oxide mixers were reported to have delivered higher levels of nitrous oxide than specification during routine testing.
Source facts
- Action
- CAREstream Medical Ltd. notified its customers on or about 05/15/2015 via a Medical Device Product Correction letter. The letter informed customers to return their units, (model number: CMI-0100-PNX-xxx) for inspection and upgrade at the firms service center in Florida. So as to allow the customer's continued use of the devices safely during this process (and until all units are upgraded), CAREsteam recommends the device only be used with an oxygen monitor with alarms, attached to the device in accordance with our product manual (refer to Users Manual (revision 6) for instructions). Customers were also instructed to fax or email CAREstream Medical Ltd. to acknowledge receipt of the Medical Device Product Correction letter.
- Initiated
- 2018-05-15
- Root cause
- Under Investigation by firm
- Terminated
- 2025-12-17
- Distribution
- Domestic Distribution: AK, AL, AR ,AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV. International Distribution: Canada and Morocco.
- Recall number
- Z-2887-2018
- Recall status
- Terminated