Skip to content

Public FDA record

Z-2887-2020: Cios Alpha, Image-Intensified Fluoroscopic X-Ray System The Cios Alpha and Cios Spin are a mobile X-Ray

Recall record Z-2887-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2887-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Aug 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens has become aware of a hardware error which affects Cios Spin and Cios Alpha systems after the previous repair of the X-ray Generator with an integrated Energy Storage Unit (ESU).

Source facts

Action
Siemens sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated July 27, 2020 to customers. This letter explains the issue that may arise while using the system. The letter instructs users to NOT disconnect the main unit from the monitor trolley when the system displays the system error message Err 16305 / 80: Confirm this error and repeat your last action to avoid potential electrical shock until pending hardware update has been completed. Our service organization will contact you for an appointment to perform the corrective action. Please feel free to contact our service organization at 1-800-888-7436 for an earlier appointment.This letter asks users to acknowledge receipt of the consignee letter by signing and returning the document. If you have any further questions, call 610-219-4834 or rebecca.tudor@siemens-healthineers.com.
Initiated
2020-07-30
Root cause
Device Design
Terminated
2021-05-20
Distribution
US Distribution to states of: NC, NE, IL, TX, OH, FL, WI, MN, MD, SC, CA, NY, VA, OK and AZ.
Recall number
Z-2887-2020
Recall status
Terminated
CDRH record ID
183128
FDA record created
2020-08-26
Recall event number
86166

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗