Skip to content

Public FDA record

Z-2887-2026: HardyDisk AST Cefiderocol 30g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30g (FDC30) for in vitro agar

Recall record Z-2887-2026 for Hardy Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2887-2026
Company in FDA record
Hardy Diagnostics
FDA classification date
Aug 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.

Source facts

Action
On 06/08/2026 an "URGENT: Medical Device Recall Notification" was sent via First Class Mail and Email and/or Facsimile to customer informing them that as a result of receiving a customer complaint/report, it was confirmed that antibiotic disk cartridges labeled as HardyDisk AST Cefiderocol (disk code FDC30) were found to contain sporadic disks labeled with Piperacillin/Tazobactam disk code (TZP110). Customers are instructed to: 1. Discard any remaining stock. 2. Response to this notification is requested within 5 business days of receipt of this letter, either through mail or email. To respond by mail, complete, sign, and send the enclosed inventory reconciliation form (FORM-2023C) via USPS First Class Mail. To respond by email, scan and send the completed inventory reconciliation form (FORM-2023C) to TechnicalServices@HardyDiagnostics.com. 3. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to invoice upon request. For Questions - contact Technical Services Department at (800) 266-2222, option 2 or via email at TechnicalServices@HardyDiagnostics.com.
Initiated
2026-06-08
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of AZ, CA, FL, ID, IN, MA, MS, NJ, NV, NY, OH, PA, and TN.
Recall number
Z-2887-2026
Recall status
Open, Classified
CDRH record ID
221095
FDA classified
2026-08-05
Recall event number
99226

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗