Open original FDA record ↗Recall record
- FDA identifier
- Z-2887-2026
- Company in FDA record
- Hardy Diagnostics
- FDA classification date
- Aug 5, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.
Source facts
- Action
- On 06/08/2026 an "URGENT: Medical Device Recall Notification" was sent via First Class Mail and Email and/or Facsimile to customer informing them that as a result of receiving a customer complaint/report, it was confirmed that antibiotic disk cartridges labeled as HardyDisk AST Cefiderocol (disk code FDC30) were found to contain sporadic disks labeled with Piperacillin/Tazobactam disk code (TZP110). Customers are instructed to: 1. Discard any remaining stock. 2. Response to this notification is requested within 5 business days of receipt of this letter, either through mail or email. To respond by mail, complete, sign, and send the enclosed inventory reconciliation form (FORM-2023C) via USPS First Class Mail. To respond by email, scan and send the completed inventory reconciliation form (FORM-2023C) to TechnicalServices@HardyDiagnostics.com. 3. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to invoice upon request. For Questions - contact Technical Services Department at (800) 266-2222, option 2 or via email at TechnicalServices@HardyDiagnostics.com.
- Initiated
- 2026-06-08
- Root cause
- Under Investigation by firm
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, FL, ID, IN, MA, MS, NJ, NV, NY, OH, PA, and TN.
- Recall number
- Z-2887-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221095
- FDA classified
- 2026-08-05
- Recall event number
- 99226