Z-2888-2016: Spiegelberg Brain Pressure Monitoring System Probe 3PN, REF: SNDI3.1.531FV534P and Probe 3PS, REF: SNDI3.

Recall record Z-2888-2016 for Spiegelberg Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2888-2016
Company in FDA record
Spiegelberg Gmbh & Co. KG
FDA record creation date
Sep 26, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg. The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.

Source facts

Action
Spiegelberg Gmbh & Co. KG sent an Urgent Medical Device Recall letter dated December 29, 2009, to all affected customers. Customers were instructed to quarantine and return stored products. Furthermore, they were asked to inform all their customers which have received the affected products and recall the products accordingly. Furthermore, they were asked to tell us about the actions taken by them.
Initiated
2009-12-29
Root cause
Packaging
Terminated
2016-10-05
Distribution
Nationwide Distribution to PA only
Recall number
Z-2888-2016
Recall status
Terminated
CDRH record ID
149491
FDA record created
2016-09-26
Recall event number
75160

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗