In the vast majority of products specified, a shrinkage of the air chamber during sterilization has occurred. Because of these false high readings were displayed in the lower measuring range from 0 to 20 mmHg.
The observed measurement error was particularly at the lowest end of the range (to 0 mmHg) up to 8 mmHg.
Source facts
Action
Spiegelberg Gmbh & Co. KG sent an Urgent Medical Device Recall letter dated December 29, 2009, to all affected customers. Customers were instructed to quarantine and return stored products.
Furthermore, they were asked to inform all their customers which have received the affected products and recall the products accordingly. Furthermore, they were asked to tell us about the actions taken by them.
Initiated
2009-12-29
Root cause
Packaging
Terminated
2016-10-05
Distribution
Nationwide Distribution to PA only
Recall number
Z-2888-2016
Recall status
Terminated
CDRH record ID
149491
FDA record created
2016-09-26
Recall event number
75160