Z-2888-2026: REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instrum

Recall record Z-2888-2026 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2888-2026
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Aug 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

Source facts

Action
Between the dates of June 26 -29, 2026, Beckman Coulter issued a Urgent Medical Device via Email and US Mail. Beckman Coulter ask consignees to take the following actions: 1. Manually verify results from the panels and assays listed below for Daptomycin or Cefepime when read with the WalkAway instrument or autoSCAN-4 instrument and for Vancomycin when read with the autoSCAN-4 instrument. 2. Continue routine quality control procedures, review flagged or unusual panel results and follow laboratory verification practices prior to reporting results. 3. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 4. Beckman Coulter recommends posting this letter on or near the affected analyzers.
Initiated
2026-06-26
Root cause
Software Design Change
Distribution
US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.
Recall number
Z-2888-2026
Recall status
Open, Classified
CDRH record ID
221385
FDA classified
2026-08-05
Recall event number
99349

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗