- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software contained three additional set configurations which caused the system to reinitialized the AST set carrier defaults.
Source facts
- Action
- BD Diagnostic Systems verbally notified all affect customers via the telephone on March 28, 2011. The notification identified the product, the problem, and the action to be taken by the customer. An "URGENT PRODUCT RECALL" follow-up letter was mailed to all customers on April 8, 2011 via UPS. Customers were instructed to complete and fax the attached reply form to acknowledge their receipt of the notification to Regulatory Compliance, 410-316-4256. The letter instructs customers to refer to the modified release notes and IMPORTANT NOTICES outlined in the attached notification letter. If further assistance is needed regarding replacements and all other inquires, please contact BD Technical Services and Support at 1-800-638-8663.
- Initiated
- 2011-03-28
- Root cause
- Software design (manufacturing process)
- Terminated
- 2011-07-29
- Distribution
- Nationwide Distribution (USA) including states of: AK, AL, AR, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
- Recall number
- Z-2890-2011
- Recall status
- Terminated
- CDRH record ID
- 100877
- FDA classified
- 2011-07-28
- Recall event number
- 59036