Z-2890-2016: The EKOS cart has the catalog number 700-60101. The EKOS cart is a custom infusion stand/cart used to ho

Recall record Z-2890-2016 for EKOS Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2890-2016
Company in FDA record
EKOS Corporation
FDA classification date
Sep 27, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Casters on the EKOS carts may be missing washers which can cause premature wear and dislodgement of the casters/wheels leading to failure.

Source facts

Action
The firm, EKOS, sent an "URGENT FIELD SAFETY ACTION" letter dated August 22, 2016 to its customers. The letters were sent to customers informing them of the field safety action, how to identify the affected carts at their facility and instructing customers locate the cart, quarantine it and do not use the cart for any procedures. A representative from EKOS Corporation will be contacting customers in the near future to arrange a date and time when the cart at a facility can be inspected and any affected caster(s) replaced. As the cart is inspected, it will be tagged to easily identify it and confirm it has been inspected and repaired as necessary. Customers will be provided with a copy of the inspection/repair form for the cart they inspect. Please retain that form for records. If customers have any questions, please contact EKOS Customer Service at 1-888-400-3567 or your local EKOS representative.
Initiated
2016-08-22
Root cause
Device Design
Terminated
2016-12-30
Distribution
US Nationwide Distribution in states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WY.
Recall number
Z-2890-2016
Recall status
Terminated
CDRH record ID
149355
FDA classified
2016-09-27
Recall event number
75100

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗