Z-2890-2026: Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

Recall record Z-2890-2026 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2890-2026
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Aug 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.

Source facts

Action
On July 6, 2026, Siemens Healthineers issued a Urgent Medical Device Recall to affected consignees via Fed Ex. Siemens Healthineers ask consignees to take the following actions: 1. Discontinue Atellica CH Urine Albumin (UAlb) testing and transition to the Atellica CH Microalbumin_2 (ALB_2) reagent, SMN 11097610, and Atellica CH Microalbumin_2 Calibrator (ALB_2 CAL), SMN 11099435, as an alternate test solution. 2. Replace all reaction cuvette segments following discontinuation of UAlb testing. Refer to the Atellica Solution Online Help for instructions on replacing reaction cuvette segments. 3. Complete the Field Correction Effectiveness Check Form attached to this letter and return within 30 days. 4.Credit will be issued for unexpired UAlb reagent (SMN 11537225), UAlb calibrator (SMN 11537247), and reaction cuvettes segments, as applicable. 5. Please review this letter with your Medical Director to determine the appropriate course of action, including whether review of any previously generated results is warranted. 6. Please retain this letter with your laboratory records and forward this letter to those who may have received the affected product.
Initiated
2026-07-06
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Curacao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿
Recall number
Z-2890-2026
Recall status
Open, Classified
CDRH record ID
221458
FDA classified
2026-08-05
Recall event number
99389

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗