Z-2892-2011: Aquamantys 2.3 Bipolar Sealer Model #23-113-1. Product Usage: The Aquamantys Bipolar Sealer is a sterile

Recall record Z-2892-2011 for Salient Surgical Technologies, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2892-2011
Company in FDA record
Salient Surgical Technologies, Inc
FDA classification date
Jul 28, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Insulating material on the shaft of the device may separate and expose the electrical conductor and result in burning the patient skin

Source facts

Action
Salient Surgical Technologies, Inc. sent an "URGENT MEDICAL DEVICE RECALL" notification letter dated July 18, 2011 to all affected customers and distributors. The letter describes the product, problem and the actions to be taken. The letter requested that customers and distributors immediately identify and discontinue use of all Aquamantys 2.3 Bipolar Sealer devices in inventory and quarantine them until they can be returned to Salient. The letter states that a Salient sales representative will arrange return of the recalled product to Salient. The letter provides information for customers about the replacement of the recalled product. If you have any questions, please contact Customer Service, at 866-777-9400, or your sales representative.
Initiated
2011-07-18
Root cause
Nonconforming Material/Component
Terminated
2012-03-09
Distribution
Worldwide Distribution - USA (nationwide) and countries of China, Denmark, France, Germany, Greece, Italy, Japan, , South Africa, Spain, and Turkey.
Recall number
Z-2892-2011
Recall status
Terminated
CDRH record ID
102436
FDA classified
2011-07-28
Recall event number
59377

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗