Z-2894-2020: Flexible DCI Intubation Fiberscope, Part: 11301BND1, with Instruction Manual: Z18445US - Product Usage: a

Recall record Z-2894-2020 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2894-2020
Company in FDA record
Karl Storz Endoscopy
FDA record creation date
Aug 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

Source facts

Action
On 07/27/2020, Urgent Medical Device Correction notices were mailed to customers. Customers were informed that this labeling change corrects the instructions for sterilization of the impacted endoscopes using STERRAD or EtO modalities. Do not use any impacted endoscope until you have confirmed that it has been appropriately sterilized in accordance with the instructions in the correction notice or via another method specified in instructions for use with updated sterilization information. When sterilizing one of the impacted flexible intubation endoscopes with the STERRAD NX Advanced Cycle or STERRAD 100NX Flex Cycle, one of the sterilization trays or containers specified below should be used: Model/Use Tray or Container: 11301AA1, 11301BND1/ASP Aptimax Tray (13837); 11301BNX, 11302BDX, 11303BNX/ASP Aptimax Tray (13837), or KARL STORZ SteriTite Container (KSZ-39406AS); 11301BN1, 11302BD2, 11302BDD2/ ASP Aptimax Tray (13837), or KARL STORZ SteriTite Container (KSZ-39402AS) The recommendation for ethylene oxide sterilization is being removed from the labeling of affected endoscopes. Customers are asked to do the following: -Examine your inventory. -Complete, sign, and return the Acknowledgement Response Form. -If impacted endoscopes are in inventory, ensure all relevant personnel in your organization are made aware of the information in this notice. -If you previously purchased a KARL STORZ plastic (Udel) sterilization trays are being used to sterilize impacted endoscopes, you may be entitled to a refund or credit. To inquire about refund/credit or for ordering assistance for SteriTite container for STERRAD sterilization, call Customer Support: 800-421-0837, press 1. -Instruction manuals are being updated for the impacted endoscopes. -Updated documents can be accessed electronically on or after August 7th at https://spwebspace.karlstorz.com/sites/ISRU. Correction Inquiries: Contact Medical Affairs Department at 800-649-8852 or medicalaffairs@karlsto
Initiated
2020-07-27
Root cause
Device Design
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.
Recall number
Z-2894-2020
Recall status
Open, Classified
CDRH record ID
182448
FDA record created
2020-08-27
Recall event number
86009

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗