Skip to content

Public FDA record

Z-2894-2026: 5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905

Recall record Z-2894-2026 for Aniara Diagnostica LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2894-2026
Company in FDA record
Aniara Diagnostica LLC
FDA classification date
Aug 5, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.

Source facts

Action
Email notifications were sent to customers on 6/17/26. Please reply to this email as acknowledgment you have received this initial notification and please include the quantity of each product which is remaining in your possession. A notification letter dated 7/1/26 was sent to customers. Actions to be implemented: " Destroy all stock of 5-PLASMA Bivalirudin Control, 5-PLASMA Bivalirudin Calibrator, & 5-PLASMA AMR Verification Kits - Please permanently deface the box labels and vial labels before disposal. " Complete and return the Awareness Acknowledgement form. " If product has been consumed please add brief statement to the Comment section, complete the End User Information, and return the Awareness Acknowledgement form.
Initiated
2026-07-01
Root cause
Process control
Distribution
US Nationwide distribution in the states of IL, NC, TN.
Recall number
Z-2894-2026
Recall status
Open, Classified
CDRH record ID
221367
FDA classified
2026-08-05
Recall event number
99344

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗