Open original FDA record ↗Recall record
- FDA identifier
- Z-2897-2011
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Jul 29, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Toshiba America Medical Systems, Inc. is initiating a field correction on the Aplio Artida Ultrasound Diagnostic System because the panel operation for the Aplio Artida may become disabled (i.e., system may hang up) when processing to display patient information (heart rate, time, VCP counter, ect.).
Source facts
- Action
- Toshiba America Medical Systems, Inc. sent out an "Urgent Medical Device Correction" dated December 21, 2010 to all affected customers. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were notified that Toshiba will issue a field modification to correct the software, and if the problem occurs, they were instructed to shutdown the system holding down the power supply button for 5 seconds for 5 seconds and then restart the system. A form to send back to Toshiba is included. For questions on this recall call (800) 421-1968 or contact your local Toshiba Representative at (800) 521-1968.
- Initiated
- 2010-10-01
- Root cause
- Software design
- Terminated
- 2012-04-20
- Distribution
- Worldwide Distribution: (USA) Nationwide distribution including the states of MD, OH, PA, NY, MA, GA, AL, TX, IL, WI, IA, MN, KY, MT, OR, LA, and CA; and the country of Puerto Rico.
- Recall number
- Z-2897-2011
- Recall status
- Terminated
- CDRH record ID
- 97589
- FDA classified
- 2011-07-29
- Recall event number
- 57856