Z-2898-2011: AxioBionics Wearable Therapy Upper Extremity BioSleeve (Right and Left) that includes 2-16 BioGel Velcro

Recall record Z-2898-2011 for Axiobionics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2898-2011
Company in FDA record
Axiobionics, LLC
FDA classification date
Aug 1, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Premature Device Failure : Failure of the electrode sites to conduct and provide muscle stimulation due to silver fabric component failure.

Source facts

Action
Axiobionics, LLC is notifying all affected customers by phone. Three attempts will be made to contact the customer by phone. Contact by telephone will be made to prescribing physician to determine if Axiobionics has latest patient contact information. If information is different, new contact information will be recorded and 3 additional attempts will be made with the new information. If the patient cannot be reached by telephone, a certified letter will be sent to the patient's address. If contacting by mail, the letter template will be provided to the FDA recall office . All patients should have been contacted by telephone, certified mail, or in person by June 10, 2011. All patients affected by the field correction will be asked the following questions to determine if their electrodes are functional and resistances are not exceeding the output capabilities of the stimulator they are using. 1. Are you using your garment? 2. Is your garment delivering the stimulation required for your therapy? 3. If the patient is using an NP4 stimulator: What are the resistance readings on each channel? 4. If the patient answers No to question 2 or states that the resistance is reading greater than 200 ohms on at least one channel in question 3, the garments will be returned to Axiobionics for repair. Garments that are not functioning correctly will be repaired first. Functioning garments will be tested by immersion into a salt bath under the given protocol to determine if the silver fabric is properly coated. Questions regarding this correction should be directed to the Axiobionics representative at 800-552-3539.
Initiated
2011-05-27
Root cause
Nonconforming Material/Component
Terminated
2013-08-29
Distribution
Nationwide Distribution--USA (nationwide) including the states of AZ, GA, IL, IN, MI, MO, NY, TX, and WY.
Recall number
Z-2898-2011
Recall status
Terminated
CDRH record ID
100917
FDA classified
2011-08-01
Recall event number
59046

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗