Skip to content

Public FDA record

Z-2899-2020: RayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as wel

Recall record Z-2899-2020 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2899-2020
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Aug 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image selected. When starting the task, an error message might be displayed stating: Review files not available.

Source facts

Action
Only 1 effected consignee who was emailed a field safety notice on 08/05/2020. This letter instructs the user to: Be aware that registration of a treatment image with a corresponding study with multiple series in the same frame of reference might block the offline image review. Verify the correctness of the data in RayCare PACS before using the data for any decision making. -Please educate staff and all users about this workaround. -Inspect your product and identify all installed units with the above software version number(s). -Confirm you have read and understood this notice by replying to the notification email.
Initiated
2020-07-29
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution including in the state of Tennessee.
Recall number
Z-2899-2020
Recall status
Open, Classified
CDRH record ID
183197
FDA record created
2020-08-27
Recall event number
86249

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.