Open original FDA record ↗Recall record
- FDA identifier
- Z-2900-2016
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Sep 28, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter is recalling the AQUIOS CL System because the use of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe.
Source facts
- Action
- An Urgent Medical Device Recall letter dated 9/6/16 was sent to customers to inform them that the use of Positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. The letter informs the customers of the impact of the recall, action to be taken, and the resolution by Beckman Coulter. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding this notice, are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support; call 1-800-369-0333 in the US and Canada; outside the US and Canada, contact their local Beckman Coulter Representative.
- Initiated
- 2016-09-06
- Root cause
- Software Design Change
- Terminated
- 2017-08-07
- Distribution
- US: IL, PA, NY, GA, WA, FL, SC. Affected geographies: Angola, Australia, Belgium, Botswana, Canada, Chile, Columbia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Israel, Italy, Malaysia, Nambia, Netherlands, Panama, Poland, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Uganda, United Kingdom and Zambia
- Recall number
- Z-2900-2016
- Recall status
- Terminated
- CDRH record ID
- 149377
- FDA classified
- 2016-09-28
- Recall event number
- 75111