Open original FDA record ↗Recall record
- FDA identifier
- Z-2900-2026
- Company in FDA record
- Instrumentation Laboratory
- FDA classification date
- Aug 6, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Source facts
- Action
- On 5/21/26, the firm began notifying consignees in the US via letter. Consignees were informed that Werfen will be proactively shipping replacement PAKs to each facility. Customers were instructed to check inventory for affected product. They should discontinue use of the affected lots and destroy any inventory on hand. Destruction should be documented via the response form provided to the customer and returned to the recalling firm. Customers should forward the notification to all affected locations at each facility. On May 28, 2026, the field action was expanded to include affected customers in China. On June 23, 2026, the field action was expanded to include additional affected lots in Italy and Germany.
- Initiated
- 2026-05-21
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
- Recall number
- Z-2900-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220892
- FDA classified
- 2026-08-06
- Recall event number
- 99128