Z-2906-2020: Cobra x5 2-Ch EMG ET Tube, 6.0mm, REF: LTE700DCS-5, Rx Only, Sterile EO UDI: B006LTE700DCS52/$$7062618C

Recall record Z-2906-2020 for Neurovision Medical Products Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2906-2020
Company in FDA record
Neurovision Medical Products Inc
FDA record creation date
Aug 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Source facts

Action
Urgent Medical Device Recall Notice, dated July 17, 2020 was sent via email on 7/21/20 and a hard copy via USPS first class mail on 7/28/20-7/29/20 to consignees advising them that Neurovision Medical Products (hereafter NMP) is voluntarily recalling Cobra 1 & 2-Channel EMG Endotracheal Tube, 6 mm, 7 mm, and 8 mm which is used for continuous EMG monitoring and status assessment of the nerves. NMP has initiated this removal due to the potential degradation of the electrode insulation material sooner than the established shelf life. This could make the electrode more susceptible to damage by dental scratching or acute bending. This may interfere with the EMG performance of the device. Failure of the device could result in a delay in the surgical procedure. Malfunction during clinical use should be evident to the clinical user by intermittent or loss of EMG signal and high electrode impedance values. We are aware of one product failure and one complaint associated with the problem. It advises customers to immediately take the following actions: Review the enclosed list of affected lots and product codes. 2. Quarantine product subject to this notice. 3. Cease distribution and use of any product from the affected lots. 4. Advise all applicable departments (including Anesthesia, OR Leadership and Materials Management). 5. Complete and return the enclosed response form as soon as possible: a. Email: customer_service@neurovisionmedical.com b. Fax: (877) 330-1727 6. NMP Customer Service will respond with a return authorization (RMA) number and return shipping label for any stock of affected lots that you may have. 7. If you have distributed this product to another location or entity, notify them of this product recall (include a copy of this notice), and contact NMP Customer Service for a return authorization (RMA). The letter stated the firm is working quickly and closely with their manufacturing team to address the issue. In the meantime, they do have an alt
Initiated
2020-07-21
Root cause
Under Investigation by firm
Terminated
2021-04-30
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.
Recall number
Z-2906-2020
Recall status
Terminated
CDRH record ID
182867
FDA record created
2020-08-27
Recall event number
86102

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗