Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Source facts
Action
On June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person.
Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.
Initiated
2026-06-15
Root cause
Incorrect or no expiration date
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Recall number
Z-2907-2026
Recall status
Open, Classified
CDRH record ID
221376
FDA classified
2026-08-06
Recall event number
99345