Skip to content

Public FDA record

Z-2910-2020: PC Unit Front Case with Keypad Replacement Kits: TC10008389 ASSY CASE FRONT W/KEYPAD 8015LS, TC10010217 A

Recall record Z-2910-2020 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2910-2020
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Sep 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.

Source facts

Action
On 08/04/20, Urgent Medical Device Recall notices were mailed to Clinical Users: If the PC unit keypad becomes unresponsive or a key gets stuck, clinicians should remove the device from service and send to Biomedical Engineering. If administering a critical medication, continue the infusion until it is safe to replace the PC unit. Cleaning personnel can minimize the potential for fluid ingress during cleaning by 1) not using a cloth that drips, 2) wringing out the cleaning cloth to squeeze out excess fluid, and 3) not spraying fluids directly onto the device. Biomedical Engineering: If an affected PC unit has an unresponsive keypad or stuck key, remove the unit from service and call the recalling firm for information replacement parts. Until 510(k) clearance, the affected PC unit keypad design will continue to be provided for replacement or repair. Inspect all PC unit keypads for bubbling, cracks, damage, layer separations, and holes. Replace the front case keypad if it fails inspection. Return the completed Customer Response Form. Provide a copy of this letter and enclosures to customers who are currently renting affected devices. Firm is currently working on a design change for the PC unit keypad and will notify customers when replacement keypads are available. Additional Information: Customer Advocacy, 888-812-3266, customerfeedback@bd.com; Training Resources www.bd.com/alaris-system-hardware-recall; Recall Support Center, 888-562-6018, SupportCenter@bd.com; Technical Support, 888-812-3229, DL-US-INF-TechSupport@bd.com; Customer Order Management, 1-800-482-4822, GMB-CTS-CustCareInfusion@bd.com On 12/16/22, Medical Device Notifications were sent to customers informing them that they may have received impacted parts for repair purposes or received product with impacted parts. Customers were provided with links to the original recall notices and asked to review the information and complete the customer response form.
Initiated
2020-08-04
Root cause
Reprocessing Controls
Distribution
U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA
Recall number
Z-2910-2020
Recall status
Open, Classified
CDRH record ID
182708
FDA classified
2020-09-04
Recall event number
85900

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.