Skip to content

Public FDA record

Z-2910-2026: Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF

Recall record Z-2910-2026 for Freudenberg Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2910-2026
Company in FDA record
Freudenberg Medical, LLC
FDA classification date
Aug 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

Source facts

Action
On June 26, 2026, Inhealth Technologies issued a Urgent Medical Device Recall Notification via FedEx to affected consignees. Inhealth ask consignees to take the following actions: 1. Do not use, ship, or distribute affected products 2. Identify and isolate all affected inventory 3. Destroy all affected devices 4. Complete the attached response form 5. Return confirmation no later than: July 8, 2026 6. Email confirmation of product destruction to: regulatoryaffairs@freudenbergmedical.com
Initiated
2026-06-17
Root cause
Labeling mix-ups
Distribution
US Nationwide distribution in the states of IN, OH, AZ, WA, FL, MO, KY, IN, NY and the countries of GBR, FRA.
Recall number
Z-2910-2026
Recall status
Open, Classified
CDRH record ID
221229
FDA classified
2026-08-06
Recall event number
99299

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗