Skip to content

Public FDA record

Z-2911-2020: Plastic Cannula HOLDEX Item # 450216 - Product Usage: - Product Usage: It is intended for use in routine

Recall record Z-2911-2020 for C.A. Greiner & Sohne Gesellschaftmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2911-2020
Company in FDA record
C.A. Greiner & Sohne Gesellschaftmbh
FDA record creation date
Aug 28, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sample leaking from the sleeve covering the needle inside the cannula

Source facts

Action
On July 27, 2020 Greiner issued a product recall notification to their distributors. It states the following actions to be taken: 1. Stop distributing the product (concerned item/lot) immediately and isolate defective products in your facility. Complete the attached Product Disposition form and fax it to Greiner at 800.726.0052 or email to patech@gbo.com. Greiner will contact you for the return or destruction of the product after receiving the attached Product Disposition Site Confirmation form. Greiner will also use this information to send you replacement stock or issue a credit. 2. Greiner requesting to provide them with a list of your customers that have received the item with this specific lot number. include: Facility, Address, Key contact, Phone number, Fax, and Date shipped. If you have additional questions, please call Technical Service at 800-515-8112. Actions to be taken by Laboratory Director, Recall Coordinator or Medical Safety Officer: 1. Stop using the product (concerned item/lot) immediately and isolate defective products in your facility. 2. Assess risks and consequences for the use of this defective product in accordance with your procedures and take appropriate action. 3. Complete theProduct Disposition Site Confirmation form and fax to Greiner Bio-One North America, Inc. at 800.726.0052 or email to patech@gbo.com. This form is to confirm that you have discarded/destroyed all products from this item/lot. We will replace the product after the completed form is sent to Greiner. If you have additional questions, please call Technical Service at 800-515-8112.
Initiated
2020-07-27
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of UT, TN.
Recall number
Z-2911-2020
Recall status
Open, Classified
CDRH record ID
182766
FDA record created
2020-08-28
Recall event number
86128

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗