Open original FDA record ↗Recall record
- FDA identifier
- Z-2911-2026
- Company in FDA record
- Contego Medical, Inc
- FDA classification date
- Aug 6, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.
Source facts
- Action
- On 6/29/2026, recall notices started being delivered to healthcare professionals and risk managers by Medtronic sales representatives who informed of the following: 1) Sales representatives will visit, conduct inventory counts, quarantine, and remove unused affected devices. 2) If replacements are desired, they can be provided. 3) Complete and return the acknowledgment form via email to recall@contegomedical.com Questions can be directed to the firm at recall@contegomedical.com and 919-459-7250 Extension 4.
- Initiated
- 2026-06-29
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of IL, FL, NC, WI, KY, AL, VA, DE, NJ, PA, IN, MI, CA.
- Recall number
- Z-2911-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221400
- FDA classified
- 2026-08-06
- Recall event number
- 99353