Z-2913-2026: navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2

Recall record Z-2913-2026 for Ventana Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2913-2026
Company in FDA record
Ventana Medical Systems, Inc.
FDA classification date
Aug 7, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window

Source facts

Action
On 05/01/2026, the firm sent via UPS Ground an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that Roche eceived two complaints outside of the U.S. concerning the navify Digital Pathology (nDP) image management system regarding an issue where images may be mismatched or associated with incorrect specimen and slide labels within a given case. The discrepancy occurs exclusively within a subset of individual cases in which a specific nomenclature is used for specimen names and the case contains multiple specimens. Customers are instructed to: " Until navify Digital Pathology software version 2.5.2 goes live on May 8th at 6:30 PM PDT (9:30 PM EDT), users should zoom into or select slide images and cross-reference specimen/block/slide labels with the correct, synchronized information in the top Case Bar, Thumbnail Panel, or Slide Info Panel in the Viewer window before proceeding with clinical analysis. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. " Complete all sections of the enclosed recall response form (TP-02782) and fax it to 1-833-237-1015 or email it to roche4435@sedgwick.com. " Distribute this UMDC within your facility and network to ensure relevant personnel are aware of the issue. " File this Urgent Medical Device Correction (UMDC) for future reference. For Questions - contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336
Initiated
2026-04-20
Root cause
Software design
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, WI, and WV. The countries of Austria, Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Romania, Spain, and Switzerland, Argentina, Bosnia & Herzegovina, Brazil, Chile, Colombia, Dominican Republic, Ecuador, Jordan, Mexico, Myanmar, Peru, Turkey, United Kingdom, Australia, Bahrain, Colombia, Hong Kong, Portugal and India.
Recall number
Z-2913-2026
Recall status
Open, Classified
CDRH record ID
220432
FDA classified
2026-08-07
Recall event number
98998

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗