Skip to content

Public FDA record

Z-2915-2016: VariaBreeze; Catalog number: 7300-0035-2. Podiatry: ablation, vaporization, incision, excision, and coa

Recall record Z-2915-2016 for New Star Lasers, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2915-2016
Company in FDA record
New Star Lasers, Inc.
FDA record creation date
Sep 30, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

FDA inspection found that customers who had received devices with promotional material with outside use indications were not notified when labeling was revised to remove these indications.

Source facts

Action
Letters sent on September 19, 2016 via US mail advised customers of the unapproved claims that may be in their literature and advised of the approved uses for each device.
Initiated
2016-09-19
Root cause
Labeling False and Misleading
Terminated
2016-12-13
Distribution
US distribution only.
Recall number
Z-2915-2016
Recall status
Terminated
CDRH record ID
149189
FDA record created
2016-09-30
Recall event number
75048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗