Skip to content

Public FDA record

Z-2916-2026: Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A

Recall record Z-2916-2026 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2916-2026
Company in FDA record
Spacelabs Healthcare, Inc.
FDA classification date
Aug 7, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to off-the-shelf display port component that has audio/video failures

Source facts

Action
On 06/04/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that the use of DisplayPort Splitter (PN 010-1975-00) may potentially result in a loss of audio, video, or both audio and video at the local Xhibit Central Station and/or extended displays. Customers are instructed to: 1. Discontinue use Discontinue use of the DisplayPort Splitter, Model number 010-1975-00, or any unapproved splitters that may be used. 2. Notify Inform all relevant personnel within your organization, including any facilities to which you have distributed the affected products. 3. Acknowledge Confirm receipt of this notification by scanning the QR code below and completing the form. Include the model number of splitter that is utilized at the facility, and if it was purchased through Spacelabs Healthcare. (PDF version of the acknowledgement form can be provided upon request) 4. Disposal Dispose of all affected components. 5. Replace Replace the affected DisplayPort Splitters following the proper installation requirements outlined in section 6 and 7 of CSN 077-0722-00. 6. Fill out the Customer Response Form using the link or QR code provided below. Spacelabs has partnered with IQVIA MedTech to coordinate the recall response process. " Link: https://iqvia-response.my.site.com/mt/fca?cid=spacelabsxhibit-av For questions or support - contact Customer Support at 1-800-522-7025 or via website https://spacelabshealthcare.com/support/
Initiated
2026-06-04
Root cause
Unknown/Undetermined by firm
Distribution
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of CANADA, FRANCE, ITALY, MALAYSIA, THAILAND.
Recall number
Z-2916-2026
Recall status
Open, Classified
CDRH record ID
221162
FDA classified
2026-08-07
Recall event number
99260

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗