Open original FDA record ↗Recall record
- FDA identifier
- Z-2917-2018
- Company in FDA record
- Elekta, Inc.
- FDA record creation date
- Aug 23, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Order Status Remains Approved When Should Indicate Complete.
Source facts
- Action
- On August 2018 the firm sent letter to all their customer with the following instructions; Recommended User Action: A user with appropriate pharmacy rights can unlock the record. This blanks out the details requiring the user to re-enter the administration details previously entered. To avoid this locking issue, close the MAR window before navigating to the CWS. This document contains important information for the continued safe and proper use of your equipment. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. Advise the appropriate personnel working with this product the content of this letter. Elekta Corrective Actions: The issue is resolved in MOSAIQ Release 2.65 Beta 02 and in release 2.64 SP10. In order to meet regulatory requirements, you are required to complete this the Acknowledgement Form and return it to Elekta immediately upon receipt but no later than 30 days.
- Initiated
- 2018-08-03
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2024-09-23
- Distribution
- AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, WA, WI, WV, and WY Australia, Canada, Ireland, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, and United Kingdom
- Recall number
- Z-2917-2018
- Recall status
- Terminated
- CDRH record ID
- 166796
- FDA record created
- 2018-08-23
- Recall event number
- 80764