Open original FDA record ↗Recall record
- FDA identifier
- Z-2921-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Aug 28, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost
Source facts
- Action
- Siemens issued Customer Safety Advisory Notice distributed dated July 24 2020 to customers via AX048/20/S. Letter states reason for recall, health risk and action to take: If the connector linking the EtCO2 module to the HemoBox should become damaged or break, the operator shall not attempt to fix the connector or re-connect the cable to the HemoBox. Only Siemens Customer Service representative or a service provider authorized by Siemens Healthineers may resolve the issue. Siemens strongly recommends that you do not attempt to fix the connector or re-connect the cable to the HemoBox until the problem has been corrected by a service engineer. Siemens is currently developing a solution to eliminate the root cause of this problem. Our service organization will contact you when a solution has been initiated to arrange a date to perform this corrective action. Please feel free to contact our service organization at 1-800-888-7436. Following the correction, the cause will be eliminated, and any recurrence of this potential fault is prevented.
- Initiated
- 2020-07-24
- Root cause
- Process design
- Terminated
- 2021-11-02
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2921-2020
- Recall status
- Terminated
- CDRH record ID
- 182968
- FDA classified
- 2020-08-28
- Recall event number
- 86179