Skip to content

Public FDA record

Z-2923-2026: Medline Micro-Kill Bleach Germicidal Bleach Wipes,7" x 8", 150 count, 150 count, Product Number MSC351400

Recall record Z-2923-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2923-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Aug 7, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medline Micro-Kill Bleach Germicidal Bleach Wipes may exhibit a reduction in bleach concentration to levels below specification prior to the end of the labeled 12-month shelf life. Reduced bleach concentration may compromise the effectiveness of disinfection, potentially allowing for the survival and persistence of pathogenic microorganisms.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/30/26 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-104 Recall Code: 3.Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Initiated
2026-06-30
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states and territories of Guam, Puerto Rico.
Recall number
Z-2923-2026
Recall status
Open, Classified
CDRH record ID
221542
FDA classified
2026-08-07
Recall event number
99432

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗