Skip to content

Public FDA record

Z-2927-2020: Dimension Vista High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in

Recall record Z-2927-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2927-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Sep 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dimension Vista High Sensitivity Troponin I (TNIH) may exhibit a negative bias with patient samples

Source facts

Action
Siemens Healthcare issued Urgent Medical Device Correction (UMDC # VC-20-03.A.US) and Urgent Field Safety Notice (UFSN # VC-20-03.A.OUS) on July 24, 2020 to Dimension Vista customers who received the Dimension Vista TNIH reagent to notify them of the issue, the potential risk to health, and actions to be taken by the customer. Customers were instructed to: "Review the letter with their Medical Director Use lot specific correlation factors with lots 20008BB, 20035BC, and 20135BB. Siemens is working to restore assay performance, until such time, subsequent lots may contain an alert card in the carton containing lot specific correlation factors ."Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. "If they have received any complaints of illness or adverse events associated with the product listed in Table 1 of the letter, immediately contact their local Siemens Remote Services Center or their local Siemens Technical Support Representative. Questions, please contact your Siemens Remote Services Center or your local Siemens Technical Support.
Initiated
2020-07-24
Root cause
Under Investigation by firm
Terminated
2023-06-07
Distribution
Nationwide
Recall number
Z-2927-2020
Recall status
Terminated
CDRH record ID
182848
FDA record created
2020-09-02
Recall event number
86151

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗