Z-2928-2026: Allura Xper FD. Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Intervent

Recall record Z-2928-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2928-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Aug 10, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firmware issue potentially leading to loss of geometry movements.

Source facts

Action
Philips notified consignees on about 06/29/2026 via certified mailed letter titled URGENT Medical Device Correction. Consignees were provided instructions to properly identify affected systems and table configurations, instructions for restarting the geometry when the issue occurs, to ensure the notification is distributed to all relevant users and retained with device documentation until corrective actions are completed. Consignees were also instructed to post the notification letter near affected devices for ease of reviewing, notify recipients if affected units were transferred, and to complete and return the provided response form. A Local Philips representative will contact consignees to schedule a visit for the implementation of the software update once available. Philips will address the identified issues by implementing a software update (R8.2.102.1) in all affected systems (FCO72200674). Philips currently expects R8.2.102.1 to be available by Q1 2027, subject to regulatory clearance.
Initiated
2026-06-29
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Lao, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Uzbekistan, Vietnam, Yemen.
Recall number
Z-2928-2026
Recall status
Open, Classified
CDRH record ID
221445
FDA classified
2026-08-10
Recall event number
99378

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗