Z-2936-2020: Parvovirus B19 IgG EIA kit Part No. V519IGUS

Recall record Z-2936-2020 for Diasorin Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2936-2020
Company in FDA record
Diasorin Inc.
FDA record creation date
Sep 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Printed label inside the box lid of the Parvovirus has the incorrect Lot Specific Constant value

Source facts

Action
DiaSorin issued a customer notification letter on July 17, 2020. The letter states the following action to be taken: Please use the Lot Specific Constant value of 0.14 with your remaining inventory of Lot 71440040AA.. The Lot Specific Constant is used to calculate the cut-off value and use of the incorrect value may lead to: 1. An invalid test due to the Low Positive Control resulting above the upper limit of the assigned range. 2. The reporting of a false positive patient result if the Low Positive Control does not fail and invalidate the test run. If you have used the affected product Lot 71440040AA in combination with the incorrect Lot Specific Constant of 0.10 and you obtained valid runs, please: 1. Review the test results of these valid runs for values between 1.1 Index and 1.54 Index, which may have been falsely reported as positive. 2. Take actions as appropriate. 3. DiaSorin recommends discussing the content of this letter with your laboratory director with respect to the need to review previous test results, conduct patient follow-up and/or repeat testing as appropriate. Complete page two of this letter and return to DiaSorin Inc. Forward this letter to anyone in your organization who may be impacted. We apologize for any inconvenience this may cause. If you have any questions regarding this notice please contact DiaSorin Inc. Technical Support at 1-800-328-1482.
Initiated
2020-07-17
Root cause
Under Investigation by firm
Distribution
NJ, WI, NY, NC, MN, CA, IL, VA, MA, MI, CT Canada
Recall number
Z-2936-2020
Recall status
Open, Classified
CDRH record ID
182850
FDA record created
2020-09-03
Recall event number
86152

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗