The use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in
intended users.
Source facts
Action
A Quality Notification dated 6/3/26 was sent to consignees via email notifying them of the recall. Consignees in the US are to coordinate replacement of devices. ROW Consignees are to scrap unopened kits in accordance with local requirements and request replacements.
Initiated
2026-05-18
Root cause
Nonconforming Material/Component
Distribution
Worldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.
Recall number
Z-2937-2026
Recall status
Open, Classified
CDRH record ID
221169
FDA classified
2026-08-12
Recall event number
99266