Open original FDA record ↗Recall record
- FDA identifier
- Z-2938-2018
- Company in FDA record
- Medical Components, Inc dba MedComp
- FDA record creation date
- Aug 25, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Source facts
- Action
- On July 12, 2018, the firm sent a Product Recall letter to its customers. Customers were provided with a full list of product codes and lot numbers and informed that those items had been recalled due to a potential for the suture wings to crack or break. Customers were informed that implanted catheters with intact suture wings do not need to be removed. However, an alternative method of catheter securement should be initiated for all implanted catheters, due to the potential for the suture wings to crack or break. Customers were instructed to return all unused recalled product to Medcomp. If product was further distributed, customers were asked to notify their sub-accounts. Contact your customer service representative for a Returned Goods Authorization (RGA) number if necessary. The phone number to call is 215-256-4201. Customers are asked to complete and return a customer response form included with the recall letter by fax (215-256-9191) or email (complaints@medcompnet.com).
- Initiated
- 2018-07-12
- Root cause
- Process control
- Terminated
- 2020-07-02
- Distribution
- Nationwide distribution; worldwide distribution.
- Recall number
- Z-2938-2018
- Recall status
- Terminated
- CDRH record ID
- 165931
- FDA record created
- 2018-08-25
- Recall event number
- 80581