Z-2938-2020: QIAstat-Dx Respiratory SARS-CoV-2 Panel REF 691223 (US IVD under FDA EUA) - Product Usage: intended for t

Recall record Z-2938-2020 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2938-2020
Company in FDA record
Qiagen Sciences LLC
FDA record creation date
Sep 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an increased rate of potential false positive results for the Influenza A (no subtype) target, which may lead to 1) inappropriate antiviral use, 2) erroneous or delayed differential diagnosis and missing of an alternative diagnosis.

Source facts

Action
On July 15, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers to inform them of the increased rate of potential false positive results for the Influenza A (no subtype) target. Customers were instructed to do the following: l -Please review and reassess all Influenza A (no subtype) positive samples tested with the affected lots. -Please note that as per Instructions for Use if only an Influenza A signal is present and no additional signal for any of the subtypes is generated, it can be due to either low concentration or, in very rare cases, a new variant or any Influenza A strain other than H1 and H3 (e.g., H5N1, which can infect humans). Such result should, therefore, be interpreted in the context of epidemiological and clinical plausibility. -If you have used the affected lot/s and you have concerns about previously generated Influenza A (no subtype) positive results on the basis of which patient treatment/management was initiated, we recommend a review of the corresponding Influenza A (no subtype) positive results. -For continued use of the affected QIAstat-Dx Respiratory SARS-CoV-2 Panel REF 691223 LOTs, Flu A (no subtype) positive results should be retested with an alternate method and interpreted in the context of current epidemiology and clinical presentation. -Review this notice with your laboratory/medical director. -IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. -Please complete Acknowledgement of Receipt attached to this letter by 29 July 2020. If you have any questions or concerns, please contact your local QIAGEN Technical Service Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Initiated
2020-07-15
Root cause
Material/Component Contamination
Terminated
2021-03-22
Distribution
US Nationwide distribution including in the states of AK, AL, CA, DC, DE, FL, GA, IL, IN, KY, MA, MD, MN, NC, NJ, NY, OH, OR, PA, TX, WA.
Recall number
Z-2938-2020
Recall status
Terminated
CDRH record ID
182845
FDA record created
2020-09-04
Recall event number
86149

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗