Z-2939-2020: BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 490

Recall record Z-2939-2020 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2939-2020
Company in FDA record
CareFusion 303, Inc.
FDA record creation date
Sep 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.

Source facts

Action
On August 4, 2020, the firm notified affected customers via mail, "Urgent Medical Device Recall", indicating the following: Actions for Clinical Users: Clinicians should remove the pump from service and send to Biomedical Engineering if the Pump Module keypad becomes unresponsive or stuck. If a critical medication is being administered, continue the infusion until it is safe to replace the Pump Module. In an urgent situation, clinicians can close the roller clamp on the IV administration set to stop an infusion. Actions for Cleaning Personnel: Follow the cleaning instructions provided in the current Directions for Use to minimize the potential for fluid ingress during cleaning. - Do not use a cloth that drips. Be sure to wring out the cleaning cloth to squeeze out excess fluid. - Do not spray fluids directly onto the device. Actions for Biomedical Engineering: Inspect all uninstalled Pump Module Door Assembly Replacement Kits noted above. All kits dated prior to January 25, 2019 are affected. If affected kits are found upon inspection, dispose per facility guidelines. An example of how to identify the date on the keypad is provided below: - Contact BD at 1-800-482-4822 to order replacement Pump Module keypad kits at no charge (either discarded by your facility or needed for remediation of the Pump Module), - Contact the BD Recall Support Center at 1-888-562-6018 to schedule remediation of the pump module at no charge if unable to perform at your location. - Maintain a list of all Pump Module serial numbers remediated to provide to BD upon completion of the activity. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and the recall instructions provided in this letter. Actions for BD Alaris System rental providers: Provide a copy of this letter to your customers who are currently renting BD Alaris System devices. Actions by BD: BD redesigned and implemented the Alaris Pump Mo
Initiated
2020-08-04
Root cause
Device Design
Distribution
Worldwide: US* (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY); United Arab Emirates (AE); Australia (AU); Canada (CA); Hungary (HU); Israel (IL); Japan (JP)* Kuwait (KW); New Zealand (NZ); Philippines (PH); Qatar (QA); Saudi Arabia (SA); Singapore (SG); Taiwan (TW); Turkey (TR)*. *There are 159 U.S. government and/or military consignees including two located in Japan (JP) and Turkey (TR).
Recall number
Z-2939-2020
Recall status
Open, Classified
CDRH record ID
182302
FDA record created
2020-09-12
Recall event number
85962

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗