During the assembly of the FRT250 cartridge assembly, an incorrect component was used. Specifically, the FRT240-HEX (2) was used instead of the FRT250-HEX.
Source facts
Action
On about 07/04/2024, Orthopediatrics notified consignees via email. Consignees were instructed to identify any affected units and spare parts within inventory, contact Customer Service, and arrange for return of affected units. Consignees were also instructed to notify all applicable personnel or customers if further distributed or transferred.
Initiated
2024-07-04
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the state of Texas and the countries of United Arab Emirates, Brazil, Canada, Netherlands.
Recall number
Z-2941-2024
Recall status
Open, Classified
CDRH record ID
209415
FDA classified
2024-08-29
Recall event number
95077