Skip to content

Public FDA record

Z-2941-2026: 60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needlel

Recall record Z-2941-2026 for BD SWITZERLAND SARL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2941-2026
Company in FDA record
BD SWITZERLAND SARL
FDA classification date
Aug 13, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Products may not have received adequate sterilization processing.

Source facts

Action
On 06/16/2026, an "URGENT: Medical Device Recall" Letter was emailed to customers informing them that, BD has identified a mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in product not being sterilized correctly. Customer are instructed to: "Immediately discontinue use of the affected lot(s). "Check all inventory locations within your facility and destroy all affected units in accordance with your facility s process. If destruction is not immediately possible at your facility, then product should be quarantined until disposal. "If product is currently in use, replace with non-impacted product as soon as possible. 1. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 2. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 3. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 4. Report any complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com For: Product Complaints, Technical Questions Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT
Initiated
2026-06-16
Root cause
Process control
Distribution
U.S.: AZ, CA, CO, FL, GA, IL, IN, KS, MD, MI, MN, MO, MS, NC, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA O.U.S.: Canada
Recall number
Z-2941-2026
Recall status
Open, Classified
CDRH record ID
221167
FDA classified
2026-08-13
Recall event number
99264

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗