Z-2942-2024: Trilogy Evo EV300 -- Model Number/ Model Description: BL2200X15B Trilogy Evo, O2, Benelux EV300; BR2200X1

Recall record Z-2942-2024 for Philips Respironics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2942-2024
Company in FDA record
Philips Respironics, Inc.
FDA classification date
Sep 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.

Source facts

Action
On July 16, 2024, the firm began notifying customers through letters titled "URGENT Medical Device Correction" - UPDATE. Customers were informed of the release of mandatory software update version 1.05.10.00 and a User Manual addendum for the Trilogy Evo, Trilogy Evo O2, Trilogy Evo Universal, and Trilogy EV300 ventilators. This letter outlines the safety issues addressed by this update, previously tracked through recalls RES 91832, 92048, 92515, 93609, and 94263. Issues include "Accuracy of Oxygen Delivery", "Environmental Contamination of Device Sensor", "Battery Depleted or Loss of Power Alarms Triggered", "Incomplete Contraindications Statement", and "Translation errors in the Korean, Traditional Chinese, and Spanish Instructions for Use manuals". Customers were instructed to immediately update the device software following instructions provided in the letter. The firm will follow up with customers to ensure devices have been updated. DME/Homecare users can access the software via the "My Philips for Professionals" website. Hospital users can access the software via In Center. If you have any questions, call 800-345-6443.
Initiated
2024-07-16
Root cause
Device Design
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: AU, BD, BE, BG, BO, BR, BS, CH, CL, NL, CN, CO, CZ, DE, DK, FR, SE, GB, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JP, KE, KW, LB, LT, MA, ME, MA, MT, MQ, MX, NO, NP, OM, PA, PE, PF, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SI, SK, SG, TH, TN, TR, TW, UA, YT, ZA and ZW.
Recall number
Z-2942-2024
Recall status
Open, Classified
CDRH record ID
209416
FDA classified
2024-09-05
Recall event number
95078

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗