Z-2944-2018: Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system

Recall record Z-2944-2018 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2944-2018
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Aug 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.

Source facts

Action
The firm, Roche, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated 7/6/18 to its customers on 7/6/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: Complete all sections of the enclosed faxback form and fax it to number 1-866-469-2231 or email it to Roche6840@stericycle.com. " Ensure you have at least a two-week supply of disposable filter tips before the Roche FES arrives to modify your system. " Provide copies of this Urgent Medical Device Correction (UMDC) to other clinicians who may need to be aware of this issue. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference. If you are using cobas p 612 pre-analytical (63x) (LCP1) systems for aliquoting in combination with non-filter tips, you must use filter tips until the root cause is identified and appropriate corrective actions have been implemented. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Initiated
2018-07-16
Root cause
Under Investigation by firm
Terminated
2019-05-16
Distribution
US Distribution to states of: CA, NY,TX, WA, and WI.
Recall number
Z-2944-2018
Recall status
Terminated
CDRH record ID
165884
FDA record created
2018-08-27
Recall event number
80617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗