Skip to content

Public FDA record

Z-2944-2020: Merge PACS

Recall record Z-2944-2020 for Merge Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2944-2020
Company in FDA record
Merge Healthcare, Inc.
FDA record creation date
Sep 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Measurements done on the Merge PACS generated MPR s may have incorrect measurements.

Source facts

Action
The firm, IBM Watson Health, sent an "Urgent Field Safety Notice" dated 05 August 2020 to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to do the following: Avoid incorrect pixel measurements by using the following workflow: Step 1: Modality acquires a series and generates reformat and sends to Merge PACS Step 2: Radiologist performs measurement on these series from the modality. Avoid black dots that may lead to additional imaging by taking any of the following actions: - Verifying study images either at modality or in another viewer. - Increasing the window level will window out the lowest pixel value (the black dots). - Viewing additional images/series within the study. Viewing of prior series/studies. - Verifying patients history, diagnosis or other clinical data. IBM Watson Health/Merge Healthcare will provide a software correction to address these issues. A representative will be in contact with you to schedule the software update. Please ensure all potential users of Merge PACS v8.1, 8.1.1 and 8.1.2 have received this notice. Please provide this notice to other organizations impacted by this action. If you have any questions regarding this notice, please contact Customer Support at WHISupport@us.ibm.com or 877-741-5369.
Initiated
2020-08-06
Root cause
Software design
Terminated
2022-11-30
Distribution
Domestic Distribution: IL, MO, NY, ND, MI, TX, FL, OH, UT, NE, MS, MN, DE, AR, CA, WI, MD, NC, NH, GA, VA, SC, PE, CT, NM, IN. International Distribution: CA, BE, UK.
Recall number
Z-2944-2020
Recall status
Terminated
CDRH record ID
183034
FDA record created
2020-09-09
Recall event number
86205

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗