Reason stated in the source
OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
Source facts
- Action
- The firm, Haemonetics, notified customers by an email titled "URGENT: MEDICAL DEVICE RECALL" on July 13, 2018 . The email described the product, problem, and actions to be taken. The customers were instructed to inspect product for defect and return units of the affected lots for replacement with non-affected units; complete the attached customer acknowledgement form via FAX TO +1-781-356-3558 OR SCAN AND E-MAIL TO: CORPORATEREGULATORY@HAEMONETICS.COM. You can contact your local customer service representative at: North America 1.800.537.2802 or email:customerservicena@haemonetics.com.
- Initiated
- 2018-07-13
- Root cause
- Process control
- Terminated
- 2020-06-16
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2950-2018
- Recall status
- Terminated
- CDRH record ID
- 166278
- FDA record created
- 2018-08-30
- Recall event number
- 80740