V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the
patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe
cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Source facts
Action
Mindray issued an "URGENT Medical Device Recall" letter on July 22, 2024. Letter states reason for recall, health risk and actions to be taken:
The V90 electronic vaporizer can continue to be used normally. If "Vaporizer Output Abnormal" alarm or other alarms associated with abnormal vaporizer output or high anesthetic agent alarms occur during use, take appropriate intervention per the A9 Operators manual and contact Mindray Technical Support for assistance.
The intervention could be switching to another V90 electronic vaporizer, or the user can decide to switch to intravenous anesthesia.
For vaporizer replacement, refer to A9 Operators manual part number H-046-017199-00, section 3.4. Vaporizer . The manual is available on the Mindray website.
Mindray s Technical Support team can assist and may be reached at (877) 913-9663
(Option 1) Monday through Friday, 8:30 a.m. 5:30 p.m. ET.
Please ensure that applicable clinical staff are aware of the content of this letter.
Mindray will replace all affected V90 electronic vaporizers located at your facility.
The V90 electronic vaporizers will be shipped directly to your facility, and a Mindray Care Team representative will contact you to schedule the installation and collect the affected V90 electronic vaporizers
Contact:
Director, Quality Operations and Regulatory Affairs
US Agent for Shenzhen Mindray Biomedical Electronics Corp., Ltd
Mindray DS USA, Inc.
201.995.8407
Initiated
2024-07-22
Root cause
Nonconforming Material/Component
Distribution
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Recall number
Z-2951-2024
Recall status
Open, Classified
CDRH record ID
209456
FDA classified
2024-09-04
Recall event number
95099