Class II

Reason stated in the source

Elevated false positive results Yersinia enterocolitica (Yersinia) from customers using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023).

Source facts

Action
The firm notified customers on or about 05/08/2018 via a letter entitled "URGENT: MEDICAL DEVICE RECALL". The Luminex informed the customer reason why they are receiving a recall notice---elevated false positive Yersinia enterocolitica (Yersinia) results when using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023). The letter described the intended use of the test kit and the distribution dates. The letter provided instructions for destroying the product if found in the inventory by a customer at a customer site. Luminex advised the customer to complete and return an enclosed Acknowledgment and Receipt form whether a customer had a recalled product on site or not. Luminex Technical Support are available assist in completing this form. Luminex Customer Experience Team contact information is as follows: 1-877-785-2323 (U.S. and Canada) (24 hours/7days) +1-512-381-4397 (Outside U.S. and Canada) (24 hours/7days) support@luminexcorp.com
Initiated
2018-05-08
Root cause
Under Investigation by firm
Terminated
2020-06-29
Distribution
Domestic Distribution: Arizona, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kansas, Michigan, Minnesota, Missouri, New Hampshire, Tennessee, Texas, Utah, West Virginia, Wisconsin.
Recall number
Z-2952-2018
Recall status
Terminated
CDRH record ID
165659
FDA record created
2018-10-11
Recall event number
80517

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗