Z-2952-2020: PrimeStore Molecular Transport Medium (PS-MTM), Model Numbers LH-1-2, LH-1-3, LH-1-4, LH-1-5, LH-1-5CC, L

Recall record Z-2952-2020 for LONGHORN VACCINES AND DIAGNOSTIC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2952-2020
Company in FDA record
LONGHORN VACCINES AND DIAGNOSTIC
FDA record creation date
Sep 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The tubes may have been delivered to some users without labels. Users may not be able to identify the tubes without these labels. If samples stored or transported in tubes are tested on certain systems or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach.

Source facts

Action
URGENT MEDICAL DEVICE CORRECTION notification letter dated 6/1/20 was distributed to the customer. Update on PrimeStore MTM (Molecular Transport Media) and guanidine thiocyanate. We are sending you this letter to notify you that PrimeStore MTM tubes may have been delivered to some users without labels. Users may not be able to identify the tubes as PrimeStore MTM tubes without these labels. If samples stored or transported in PrimeStore MTM tubes are tested on the Hologic Panther or the Hologic Fusion System or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach. The Hologic Panther and the Hologic Fusion System platforms have a built-in decontamination step post-run analysis. Currently the Hologic decontamination liquid contains bleach that creates a dangerous chemical reaction producing cyanide gas when mixed with guanidine thiocyanate, one of the components of PrimeStore MTM, lysis buffers and other sample collection devices. The same reaction may occur if the PrimeStore MTM media comes into contact with bleach or other reagents containing bleach. You should evaluate procedures in your laboratory to determine if there is a risk of the PrimeStore MTM media contacting bleach or reagents containing bleach, such as during routine cleaning or decontamination following a spill. Please process the inactivated samples in PrimeStore MTM on molecular diagnostics platforms that do not use or contain bleach, and dispose of all the waste by following your Laboratorys Waste Disposal Protocols for your routine nucleic acid extraction kits (for DNA or RNA) as the guanidine in PrimeStore MTM aligns with the percentage of guanidine in other lysis buffers. Should there be any accidental spills of inactivated samples in PrimeStore MTM tubes, please
Initiated
2020-06-03
Root cause
Error in labeling
Distribution
Nationwide distribution including in the state of MD.
Recall number
Z-2952-2020
Recall status
Open, Classified
CDRH record ID
182998
FDA record created
2020-09-12
Recall event number
86189

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗