Open original FDA record ↗Recall record
- FDA identifier
- Z-2953-2011
- Company in FDA record
- International Technidyne Corp.
- FDA classification date
- Aug 5, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Certain lots of Hemochron Jr. PT Cuvettes (J201) demonstrate a higher bias than historically observed when compared to laboratory reference instruments.
Source facts
- Action
- International Technidyne Corp. (ITC) sent "Urgent Medical Device Recall" letters dated May 10, May 25 and June 09, 2011 to all affected customers. The letter included a description of the problem, asked customers fill out the attached form, and return affected product for replacement. For questions, please contact ITC Technical Support at (800) 631-5945.
- Initiated
- 2011-05-10
- Root cause
- Device Design
- Terminated
- 2012-11-02
- Distribution
- Worldwide Distribution: (USA) Nationwide Distribution including the countries of; Austria, Australia, Canada, Switzerland, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Korea, Liechtenstien, Libya, Netherlands, Norway, Oman,Portugal, Qatar, Saudi Arabia, Sweden, Turkey, Taiwan, and South Africa.
- Recall number
- Z-2953-2011
- Recall status
- Terminated
- CDRH record ID
- 98093
- FDA classified
- 2011-08-05
- Recall event number
- 58043