Skip to content

Public FDA record

Z-2953-2018: Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc.,

Recall record Z-2953-2018 for Sysmex America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2953-2018
Company in FDA record
Sysmex America, Inc.
FDA record creation date
Aug 30, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.

Source facts

Action
The firm, Sysmex, issued a "Product Notifications" letter dated 7/2018 via U.S. mail on 7/27/2018 to the consignees who received the affected lots. The letter was also issued to all other XN-Series customers with registered email addresses with the Sysmex Customer Resource Center via email blast on 7/27/2018 who did not directly receive the affected lots but could have received them through other channels. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1. follow your internal procedures for evaluation of flagged results. 2. Distribute this Product Notification to the appropriate personnel to take the following actions: 3. Upon receipt of unaffected lots of replacement reagent, immediately exchange any affected Fluorocell PLT cartridges currently in use on your analyzer. 4. Discard any remaining affected lots in inventory according to your laboratorys protocols and procedures. 5.Please proceed to the URL: http://pages.sysmex.com/fluorocellrecall.html to complete the Fluorocell Recall Response Form. Confirm receipt and understanding of the Product Notification. Sysmex will replace the quantity of affected lot(s) identified in this Product Notification that are confirmed to be currently in your inventory. If you have any questions concerning this information, please contact Sysmex Quality Assurance Department at quality@sysmex.com.br.
Initiated
2018-07-27
Root cause
Nonconforming Material/Component
Terminated
2019-07-08
Distribution
Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Argentina, Brazil, Canada, Chile, Costa Rica, Ecuador, Honduras, and Nicaragua.
Recall number
Z-2953-2018
Recall status
Terminated
CDRH record ID
166208
FDA record created
2018-08-30
Recall event number
80716

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗